IRt/c.2/15ACF Base Model

Product Overview

The
IRt/c.2/15ACF Base Model
features

• Infrared Thermocouple IRt/c • Non-contact • Self Powered • Intrinsically Safe • Repeatability 0.02°F (0.01°C) • Interchangeability ±1% • Resolution approx. 0.0001°C • Close Focus

Technical Data

Target Surface Type
Hi E (non-metal)
Sensing Range
0 to 2500°F (-18 to 1370°C)
Target Surface Type
Hi E (non-metal)
Lo E (metal)
Sensing Range
0 to 2500°F (-18 to 1370°C)
500 to 2500°F (260 to 1370°C)
Optimum Range Selections
One model each J, K: adjustable over entire sensing range, output tables available
Minimum Spot Size at dist.
(with supplied apertures)
Minimum Spot Size
0.11 x 0.35” (2.9 x 8.7 mm) at 1.7” (43 mm) from sensor
Field-of-View at > min. spot
30° approximately
Spectral Response
2 to 20 μ
Output Impedance
Spectral Response
2 to 20 μ
0.1 to 5 μ
Output Impedance
Output Impedance
10 Kohms approx
Cable
Twisted shielded pair of base thermocouple material (J,K,etc.), 3 ft (.9 m) std length, Teflon sheathed, rated to 392°F (200°C) continuous service
Dimensions
2.02” x 1.375” Dia. (51.3 x 35 mm)
Weight
6.8 oz (192 g) with cable
Housing
Stainless steel, hermetically sealed, exceeds NEMA 4,4x; IP65,67, intrinsically safe, cable shield grounded to housing and electrically isolated from signal
Air Purge
Built-in; cooling capacity to 400°F (200°C) ambient; 3’ (0.9 m) polyurethane tubing provided

Set-up and calibration instructions

For all IRt/c Models with “A” in model designation (IRt/c.xxxA)

1. Connect air purge first if installing in process already at operating temperature.   Provide minimum 5 psig (30 kPa) air pressure.
2. Install IRt/c and align to view the desired target. Bring target to operating temperature if not already there. Connect leads to readout device to be used (controller, PLC, etc.).
3. If the target temperature is not known, measure the target temperature with an accurate reference. Remove the setscrew to expose the calibration screw. Adjust the calibration screw to obtain reading desired. Replace the setscrew cover when complete. For final process adjustments, the ZERO or OFFSET adjustments available on readout devices can be conveniently used.
• Installation and calibration complete.
• To maximize the linear range, see Tech Note #70.
• Calibration screw operates like a radio volume control: clockwise increases signal.